Mediana P10 Instrukcja Obsługi Strona 23

  • Pobierz
  • Dodaj do moich podręczników
  • Drukuj
Przeglądanie stron 22
17
b. Press the Print Button when all the parameter signals display normally.
c. Verify the parameter values and waveforms are printed out with 25 mm/s.
d. Set Print Speed to 50 mm/s.
e. Press the Print Button when all the parameter signals display normally.
f. Verify that the parameter values and waveforms are printed out with 50 mm/s.
g. Verify the two ECG waveforms are printed out.
h. Repeat this test for other selections.
8. Test #5: Print-On-Alarm
a. Set Print-On-Alarm to ON via the Alarm Limits Menu.
b. Set the Heart Rate of your ECG simulator to 30bpm.
c. Verify Low Heart Rate/Pulse Rate limits violated” alarm is activated and
the parameter values and waveforms are printed out.
Note: If no printer is installed in the monitor, Print Mode, Print Speed and Print On Alarm
will not display.
Note: If there is no printer paper left or printer paper feeds improperly, the monitor will
display an alarm message.
Przeglądanie stron 22
1 2 ... 18 19 20 21 22 23 24 25 26 27 28 ... 109 110

Komentarze do niniejszej Instrukcji

Brak uwag